Set commands
COMMAND CONCEPT MESSAGE ANSWER
SVT Set Temperature (ºC) #NNSVT<BL>+/-000,0KK<CR><LF> #NNOKKK<CR><LF> *
SVP Pressure (mbar) #NNSVP<BL>0000KK<CR><LF> #NNOKKK<CR><LF> *
RUN Run #NNRUNKK<CR><LF> #NNOKKK<CR><LF> *
RUNT Run temperature #NNRUNTKK<CR><LF> #NNOKKK<CR><LF> *
RUNP Run pressure #NNRUNPKK<CR><LF> #NNOKKK<CR><LF> *
STOP Stop #NNSTOPKK<CR><LF> #NNOKKK<CR><LF> *
STPT Stop temperature #NNSTPTKK<CR><LF> #NNOKKK<CR><LF> *
STPP Stop pressure #NNSTPPKK<CR><LF> #NNOKKK<CR><LF> *
TBY Standby time #NNTBY<BL>00KK<CR><LF> #NNOKKK<CR><LF> *
TRU Run time #NNTRU<BL>00,0KK<CR><LF> #NNOKKK<CR><LF> *
SRLK On pump #NNSRLKKK<CR><LF> #NNOKKK<CR><LF> *
RRLK Off pump #NNTRUKKK<CR><LF> #NNOKKK<CR><LF> *
INT Inhibit key pad #NNINT<BL>0KK<CR><LF> #NNOKKK<CR><LF> *
INTT Inhibit temperature key #NNINTT<BL>0KK<CR><LF> #NNOKKK<CR><LF> *
INTP Inhibit pressure key #NNINTP<BL>0KK<CR><LF> #NNOKKK<CR><LF> *