Immunity Test IEC60601-1-2
Test Level
Compliance Level
Electrostatic
Discharge (ESD)
IEC 61000-4-2
+/- 8 kV contact
+/- 2 kV , +/- 4 kV,
+/- 8 kV, +/- 15 kV air
+/- 8 kV contact
+/- 2 kV , +/- 4 kV,
+/- 8 kV, +/- 15 kV air
Radiated RF
Electromagnetic
Field Immunity
Test IEC 61000-4-3
10 V/m 80 MHz – 2.7 GHz
80% AM at 1 kHz
10 V/m 80 MHz – 2.7
GHz
80% AM at 1 kHz