ST-200 aQua Operator’s Manual
64
PATIENT ID:
Patient ID should be ON to get previous stored results. If patient –ID option is
deselected then instrument will not ask to enter patient ID while analyzing.
SETUP MENU
1.ANALYZE TYPE
2.DATE/TIME
3.PRINTER
4.PRINT-PARAMETERS
5.PRINT-mV
6.PATIENT-ID
Press 6
↓
PATIENT-ID
- - - - - - - - - - -
STATUS : OFF
- - - - - - - - - -
1→ MANUAL 2→ AUTO
N→ OFF
We can set manual & auto status for PATIENT ID (i.e.) press 2 for AUTO status, it
takes the PATIENT ID spontaneously in serial wise for each sample without giving
PATIENT ID.
In MANUAL status (i.e.) we have to provide PATIENT ID for each sample as per
requirement.
D) Parameters:
Depending upon requirement change model of the analyzer ensure that
Electrode blocks stacking should be done according to the order in
parameter option.
Where * indicates blank electrode.
Press Menu
Press 4
↓
1. K Na * * * *
2. K Na * * * Cl
3. K Na iCa * * *
4. K Na iCa * * Cl
5. K Na iCa Li * *
6. K Na iCa pH * *
Y → Next N→ Exit
MENU
1.DIAGNOSTICS
2.USER MENU
3.SETUP MENU
4.PARAMETERS
5.SELECT iCa UNIT