WARNING Operating your ECLIPSE outside the recommended operating
environment may impact performance, cause damage and will void
the warranty.
WARNING Pulse Flow settings should be determined for each patient individually.
Settings from Continuous Flow applications may not be applicable to
Pulse Flow Mode.
WARNING When operating in Pulse Flow Mode, to ensure adequate breath
detection, the length of the tubing/cannula must not exceed 7 feet
(2.1m).
WARNING As with all conserving devices, the ECLIPSE may not be able to detect
some respiratory efforts.
WARNING DO NOT change from prescribed flow rate setting without consulting
with a qualified clinician.
WARNING AVAILABILITY OF AN ALTERNATE SOURCE OF OXYGEN IS
RECOMMENDED IN CASE OF POWER OUTAGE OR MECHANICAL
FAILURE. CONSULT YOUR PHYSICIAN FOR THE TYPE OF BACK-UP
SYSTEM REQUIRED.
DO NOT IGNORE ALARMS.
W
AR
N
ING
Use of certain o
xygen administration accessories not specifi
ed for
use with this oxygen system may impair the performance.
ONLY use
recommended oxygen delivery accessories.
W
AR
NING
T
he tube fi
tting may be tight.
D
O NO
T
use oil or grease to lubricate.