Model 1000 User Manual

Questions and Answers

  • C
    Chad BoydAug 19, 2025
    What to do if SeQual Eclipse 1000 shows 'UNIT NEEDS ATTENTION - Flow Rate Error/ Blocked Flow'?
    • C
      christineandersonAug 19, 2025
      If your SeQual Oxygen Equipment displays a 'Flow Rate Error/ Blocked Flow' message, inspect the tubing and/or humidifier for any obstructions. Also, check and clean the air intake filter. If the condition continues for more than 10 minutes, contact your dealer/provider.
  • C
    Crystal SalazarAug 21, 2025
    What to do if SeQual Oxygen Equipment shows 'UNIT NEEDS ATTENTION - O2 Concentration < 85%'?
    • I
      iguerreroAug 21, 2025
      If your SeQual Oxygen Equipment displays an 'O2 Concentration < 85%' message, continue use and check or clean the air intake filter. If the condition persists, contact your dealer/provider.
  • K
    Kimberly CopelandAug 23, 2025
    What to do if SeQual Eclipse 1000 Oxygen Equipment shows 'UNIT NEEDS ATTENTION - O Concentration < 70%'?
    • J
      Jennifer BatesAug 25, 2025
      If your SeQual Oxygen Equipment displays an 'O Concentration < 70%' message, continue use and check or clean the air intake filter. If the condition persists, contact your dealer/provider.
  • J
    Joann BrownAug 26, 2025
    Why is my SeQual Oxygen Equipment not turning on?
    • C
      Christine HansenAug 26, 2025
      Your SeQual Oxygen Equipment might not be turning on for several reasons. Ensure a charged Power Cartridge is installed or plug the unit into an external power source. If the Power Cartridge is discharged, plug into external power. If the issue persists, there may be a malfunction, and you should contact your dealer/provider.
  • D
    David DominguezAug 28, 2025
    What to do if my SeQual Eclipse 1000 has no oxygen?
    • G
      Gene OconnellAug 28, 2025
      If your SeQual Oxygen Equipment isn't delivering oxygen, first check the tubing for any blockages or kinks and replace if necessary. Ensure the air inlet filter is clean. Inspect the humidifier attachment and tubing. Make sure the device is turned on. Check that the tubing, cannula, and connections are properly connected and not kinked. Install a charged Power Cartridge or connect to an external power source. If none of these steps resolve the issue, contact your dealer/provider.
  • A
    andrewhicksAug 30, 2025
    What to do if my SeQual Eclipse 1000 has shutdown due to loss of external power?
    • X
      xhorneAug 30, 2025
      If your SeQual Oxygen Equipment shuts down due to a loss of external power and you don't have a charged Power Cartridge installed, use an alternative oxygen supply until the external power is restored or install a charged Power Cartridge.
  • J
    John MclaughlinSep 1, 2025
    What to do if my SeQual Oxygen Equipment has low oxygen concentration?
    • R
      robertsjohnnySep 1, 2025
      If your SeQual Oxygen Equipment is showing a low oxygen concentration, check for any restrictions in the tubing or humidifier and repair or replace if needed. Also, inspect the air inlet filter and clean or replace it if it's restricted. If the problem persists, contact your dealer/provider.
  • R
    richard25Sep 4, 2025
    Why is there no oxygen delivered in pulse mode on my SeQual Eclipse 1000 Oxygen Equipment?
    • J
      Joseph SpenceSep 4, 2025
      If your SeQual Oxygen Equipment isn't delivering oxygen in Pulse Mode, ensure the tubing/cannula is no longer than 7 feet (2.1m). Remove any humidifier attached. If the issue persists, contact your dealer/provider.
  • D
    Dean MeyersSep 6, 2025
    What does 'UNIT NEEDS ATTENTION - Low or Warm Power Cartridge' mean for SeQual Eclipse 1000 Oxygen Equipment?
    • T
      Tara JonesSep 6, 2025
      If your SeQual Oxygen Equipment displays 'UNIT NEEDS ATTENTION - Low or Warm Power Cartridge', plug the unit into external power or replace the current cartridge with a charged one.
  • R
    riggsraymondSep 9, 2025
    What to do if my SeQual Oxygen Equipment has low oxygen flow?
    • K
      Kimberly ArnoldSep 9, 2025
      If your SeQual Oxygen Equipment has a low oxygen flow, check for any restrictions in the humidifier or tubing and repair or replace them as needed. Also, inspect the air inlet filter and clean or replace it if it's blocked. If the problem continues, contact your dealer/provider.

Summary

Important Safety Instructions

Symbols Used in User Manual and on Oxygen System

Explains symbols and warnings used in the manual and on the oxygen system.

Indications for Use

Contraindications

Lists conditions or situations where the ECLIPSE system should not be used.

Safety Guidelines

Oxygen Therapy Safety

Provides critical safety information regarding oxygen therapy.

Operating Safety

Details safe practices for operating the ECLIPSE oxygen system.

Power Cartridge Safety

Outlines safety precautions related to the power cartridge.

Understanding the Controls and Front Panel

Description of the ECLIPSE

Identifies key features and components of the ECLIPSE oxygen system.

Control Panel Features

Explains the icons, numbers, and indicators on the control panel.

User Controls

Describes the function of the control buttons and indicators.

System Status Indicators

Details the meaning of the green, yellow, and red indicators on the panel.

Recommended Operating Environments

The Proper Location

Advises on selecting an appropriate location for the ECLIPSE system.

Operating Instructions

Before Operating

Provides initial steps and important information before operating the system.

Powering Your ECLIPSE System

Explains how to power the ECLIPSE system using different sources.

Operating Your ECLIPSE

Guides the user through the steps of using the ECLIPSE system.

Cleaning, Care and Routine Maintenance

Routine Maintenance

Outlines routine maintenance tasks performed by technicians and users.

Patient Cleaning and Care

Provides instructions for cleaning and caring for the system by the patient.

SeQual Eclipse 1000 Specifications

General IconGeneral
AC Power100-240 VAC, 50/60 Hz
Operating Temperature41°F to 104°F (5°C to 40°C)
Operating Humidity10% to 95% non-condensing
Battery LifeUp to 5 hours
Pulse Dose Settings1 to 6
DC Power12 VDC
Operating AltitudeUp to 10, 000 feet (3, 048 meters)
Power Requirements100-240 VAC, 50/60 Hz