6 Adjustment
The device must be adjusted if the deviation of the indication accuracy is outside the permissible
range (see chapter 5.4).
6.1 Tools and requirements
Required tools:
z Suitable test set (e. g. SPE HG, SPE-Y)
z Test head
z Test gas, depending on the type of gas of the relevant device (see table in chapter 5.4)
Youcanndallinformationonhowtoconnectthetestsettothedeviceintheoperatinginstructions
of the test set.
Requirements for a proper adjustment:
z Device is switched off
z Test head is not yet attached
z Atmosphere of the environment is not polluted (e. g. fresh air)