7.4 Error Correction Table
Fault during startup
Display Fault Reason Effect Correc-
tion
None No dis-
play after
switch-on
Battery
dead
Hard-
ware
does not
start up.
Charge
battery
or install
new
battery.
If possi-
ble:
"internal
error"
mes-
sage
After
switch-
on, "in-
ternal
error"
message
appears.
RAM/
EPROM
error
display/
keyboard
defective
No mea-
suring
possible
Replace
BT 200.
Fault during operation
Display Fault Reason Effect Correc-
tion
Battery
display
flashes
before.
Device
goes off.
Battery is
dead.
No mea-
suring
possible
Charge
battery/
install
new
one.
None Device
goes off.
Time with
no user
activity
was
excee-
ded.
None Press
ON/
OFF
button.
Internal
driver
defect
Internal
driver is
defective.
HW de-
fect
No stati-
on/
branch
test
possible
Re-
place
BT 200.