RECOMMENDED
FLOW
RATE
ACCURACY
TEST
MATERIALS
REQUIRED:
1.
2.
4.
a
SIGMA
6006+
infusion
pump.
Fluid
Source
IV
set
of
the
brand
specified
on
the
pump's
label
(see
inside
the
door
or
top
of
case).
Note
that
if
pump
is
being
accuracy
tested
following
a
clinical
problem,
it
is
essential
to
test
with
the
exact IV
set
used
when
the
problem
occurred.
24"
IV
container
(250,
500
or
1000ml)
to
be
used
as
pumping
É
source.
(Collection
Vessel)
Collection
vessel
to
be
100ml
graduate.
_
TEST
DATA
Flow
Rate
(ml/hr)
test
being
run
at
100
Suggested
collection
vessel
100ml
graduate
Pump
Vol.
Limit
(ml)
setting
during
test
(length
of
test)
50
(30min)
Optimal
actual
volume
pumped
(ml)
50ml
Pass
Criteria:
acceptable
collected'pumping
volumes
(ml)
|
47.5ml—
52.5ml
(+
5%)
Test
Failed
Test
Passed.
PROCEDURE
1.
Hang
the
IV
container
above
the
pump.
Connect
the
IV
set
to
the
IV
container
and
prime
the
set
completely
(remove
all
air
from
the
tubing,
be
sure
the
drip
chamber
is
at
least
half
full).
Load
the
IV
set
into
SIGMA
6000+
pump
and
close
the
pump
door.
Be
sure
all
IV
set
clamps
are
open,
Connect
the
end
of
the
IV
set
to
the
collection
vessel.
If
vessel
is
a
100ml
graduate,
locate
the
end
of
the
set
at
about
the
55ml
level.
Be
sure
fluid
is
at
the
end
of
the
IV
set,
but
none
has
dripped
into
the
graduate
before
starting
the
test.
Tura
the
pump
on,
set
the
flow,
rate
=
100
(mi/hr)
and
volume
limit
=
50ml.
Begin
the
test
by
pressing
the
pumps’
RUN
key.
.
Allow
the
SIGMA
6000+
to
pump
until
it
goes
into
Infusion
Complete
alarm.
If
test
is
interrupted
by
any
alarms;
except
air,
the
test
can
be
continued
as
long
as
the
pump
was
observed
by
the
tester
to
have
not
been
in
stop
for
more
than
a
few
minutes,
otherwise
press
STOP
to
halt
the
KVO
(keep
vein
open)
rate
once
the
pump
goes
into
Infusion
Complete.
Observe
the
actual
volume
collected.
If
the
collected
volume
is
within
the
range
47.5ml
52.5ml,
the
pump
is
properly
calibrated.
If
not,
return
the
pump
to
SIGMA
for
recalibration
or
follow
SIGMA’s
Service
Procedure
Flow
Rate
Calibration.
Section
1
Page
13