Functional Specifications | Sweep Normal
26
timebase adjustments only. If other adjustments are required, the analyzer must be returned to
Signal Hound.
Contact Signal Hound for more information regarding calibration software and required
equipment, or to schedule a calibration.
6 Functional Specifications
6.1 Sweep Normal
IBW 40MHz
Frequency range 100kHz to 20GHz / 43.5GHz
RBW range 0.1Hz to 3MHz
RBW / VBW ratio 1 to 1000, selectable/arbitrary
Sweep speed 160 GHz/sec @ 10 kHz RBW
18 GHz/sec @ 1 kHz RBW
6.2 Sweep Fast
In Spike, fast sweep measurement mode is active when real-time measurement mode is selected
with a span greater than 160MHz.
IBW 160MHz
Frequency Range 100kHz to 20GHz / 43.5GHz*
RBW Range 30kHz to 10MHz
VBW Ratio 1 (VBW not selectable)
Sweep Speed 1THz/s
*Below 650MHz center frequency, the sweep speed drops below 1THz/s due to the smaller IF bandwidth
available at lower frequencies. Above 24 GHz, the sweep speed drops below 1 THz/s for certain spans
due to hardware settling time.
6.3 Real Time (40MHz 160MHz span)
IBW 160MHz
Frequency Range 700MHz to 20GHz / 43.5GHz
RBW Range 30kHz to 10MHz
VBW Ratio 1 (VBW not selectable)
6.4 Real Time (< 40MHz span)
IBW 40MHz
Frequency Range 100kHz to 20GHz / 43.5GHz

Other manuals for Signal Hound SM200A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Signal Hound SM200A and is the answer not in the manual?

Signal Hound SM200A Specifications

General IconGeneral
BrandSignal Hound
ModelSM200A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals