9.2 WATER AND FILTER CONNECTION
Temperature of the water supply
°C
30 MAX
Hardness of the water supply
°f
7 – 12
Pressure of the water supply kPa 200/400
FIX THE FILTER TO THE BRACKET CONNECT 2 INLET PIPES
FOLLOWING THE ARROW
DIRECTION: MAIN FAUCET -
FILTER – FILTER - MACHINE
ONCE PIPES ARE
FIXED, OPEN THE
WATER AND CHECK
THE TIGHTENING
10
CARBON FILTER
11
FILTER MUST BE CHANGED AT
LEAST TWICE A YEAR
12
Install the carbon or mesh filter supplied on the water supply cock. The water hardness must be between 7 and 12°F and the maximum temperature 30°C.
The filters must be replaced at least twice a year.
Failed observance of these parameters will cause the internal osmotic membrane to deteriorate rapidly.
Make sure the water hardness is between 7 and 12°F, otherwise install a water softener to prevent damages caused by scale, which are not covered by warranty.
SECURE THE BRACKET
ON THE WALL AT MIN.
50 CM FROM THE
GROUND
10/05/2019 6