114 GB
Error Problem Solution
Battery sym-
bol on the LC
display
4
Weak battery Insert a new
battery
“Lo” on the
LC display
4
(measured
value below
32.0°C or
89.6°F)
Thermometer
correctly posi-
tioned in the ear
or on the fore-
head/temple?
Pay heed to the
operating in-
structions in re-
gard to position-
ing in the ear or
on the fore-
head/temple.
“Hi” on the
LC display
4
(measured
value above
42.9°C or
109.2°F)
Check the meas-
urement mode.
Please observe
the directions
for use with re-
gard to taking
correct measure-
ments.
“ErrE” on the
LC display
4
Ambient tempera-
ture outside the
range of 15.0 °C
- 40.0 °C /
59.0 °F -
104.0 °F (fore-
head mode) and
10.0 °C -
40.0 °C /50.0 °F
- 104.0 °F (ear
and scan mode).
Please ensure an
ambient tempera-
ture in the range
of 15.0 °C -
40.0 °C /
59.0 °F -
104.0 °F (fore-
head mode) and
10.0 °C -
40.0 °C/
50.0 °F -
104.0 °F (ear
and scan mode)
Functions / error analysis
IB_96247_SSOT6B2_ES_PT_DE_GB_LB5.indd 114IB_96247_SSOT6B2_ES_PT_DE_GB_LB5.indd 114 19.11.13 16:4819.11.13 16:48