Loading...
10 GB/IE
Introduction
Temperature scale: °C (°Celsius)
°F (°Fahrenheit)
Time display: 12 / 24-hour format
Automatic switch-off : Automatic switch-off after 1 minute
Measurement range
forehead & ear mode: 32°C - 42.9°C (89.6°F - 109.2°F)
Measurement
accuracies: ± 0.2°C (± 0.4°F) in the range
from 35.5°C - 42.0°C
(95.9°F - 107.6°F)
± 0.3°C (± 0.5°F) in the range from
32.0°C - 35.4°C (89.6°F - 95.7°F)
and in the range from 42.1°C -
42.9°C (107.8°F - 109.2°F)
Measurement range
room temperature: 5.0°C - 59°C (41.0°F - 138.2°F)
Measurement
accuracy: ± 2.0°C (± 4.0°F)
Measurement
range scan mode: –22.0°C - 80.0°C
(–7.6°F - 176°F)
Measurement
accuracy: ± 2.0°C (± 4.0°F)
Ambient temperature
Ear and scan mode: +10.0 °C - +40.0 °C
(50.0 °F - 104.0 °F)
Forehead mode: +15.0 °C - +40.0 °C
(59.0 °F - 104.0 °F)