EasyManua.ls Logo

SilverFox 2246BN - Troubleshooting

SilverFox 2246BN
16 pages
Print Icon
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
If problem is still not solved after above operations,
please contact maintenance center or supplier to solve it.
Motor can’t work
normally
Abnormal Sound
1. The power cord is unconnected or
not plugged well.
1. The rotating parts are short of
lubricating oil.
2. Rotating positions (such as hinge,
POM sets, gas lift) are short of
lubricating oil.
2. Poor quality of the hand controller
or foot controller.
3. Poor quality of the control box.
4. Repture of motor tail.
5. Motor was damaged.
6. Emergency Switch has been
pressed but not reset.
3. Motor quality problems.
Connect the power cord or plug it well.
1. Add lubricating oil.
2. Add lubricating oil.
Replace the hand controller or foot
controller.
Please contact maintenance center or
supplier for troubleshooting.
Reset the emergency switch.
3. Replace the motor.
The Fault Phenomenon Reason Solution