scorpION3 5200969 Rev 4 34
4.3 System Calibration
System calibration is a complete analysis of performance and
includes any adjustments that may be necessary to restore the
system. This should be performed after maintenance and cleaning
of the emitters has been completed. Adjustment to output level
(first) and ion balance settings (second) should be made only after
the emitters have been cleaned, and after every emitter cleaning or
replacement. Adjustments must be saved using “save settings local
to bar” or they will be lost when power is removed from the bar.
Standard system tests include charge discharge time and offset
voltage. Test data should include temperature, humidity, and air
velocity. Tests of the system are made in accordance with ANSI/
ESD STM 3.1-2006 Ionization, Section 6.2 laminar flow hood
ionization. Contact Simco-Ion for information concerning test
instruments and services.