Operation
charged.
Disconnect
shunted
component
from
the
circuit
be-
fore
measuring
its
resistance.
b.
Set
the
range
switch
to
one
of
the
resistance
range
positions
as
follows
(Figure
4-8):
1.
Use
R
X
1
for
resistance
readings
from
0
to
200
ohms.
2.
Use
R
x
100
for
resistance
readings
from
200
to
20,000
ohms.
3.
Use
К.
Хх
10,000
for
resistance
readings
above
20,000
ohms.
с.
Set the
function
switch
at
either
-DC
or
+
DC
position:
Operation
is
the
same
in
either
position
except
if
there
are
semiconductors
in
the
circuit.
(See
Paragraph
4.2.1)
Adjust
ZERO
OHMS
control
for
each
resistance
range.
d.
Observe
the
reading
on
the
OHMS
scale
at
the
top
of
the
dial.
Note:
The
OHMS
scale
reads
from
right
to
left
for
increasing
values
of
resistance.
e.
To
determine
the
actual
resistance
value,
multiply
the
reading
by
the
factor
at
the
switch
position.
(K
on
the
OHMS
scale
equals
one
thousand.)
420
RESISTANCE
MEASUREMENT
OF
SEMICONDUCTORS
Make
sure
that
the
OHMS
range
being
used
will
not
damage
any
of
the
semiconductors
(refer
to
Table
1-1,
item
8,
Section
1).
4.20.1
If
there
is
a
"forward"
and
"backward"
resistance
such
as
in
diodes,
the
resistance
should
be
relatively
low
in
one
direc-
tion
(for
forward
polarity)
and
higher
in
the
opposite
direction.
4-24