OPERATION
36
5. If there is a short circuit in the DUT during the test, the red indicator light will
illuminate on the RESET switch and an alarm will sound. The display will show:
models 2925 and 2935
HI- F A I L MX XXX . X s
X . XX K V A C O F L mA
model 2945
W - H i g h MX XXX . X s
X . XX K V A C O F L mA
6. If a flash over occurs in the DUT during the test, which results in an OFL condition
due to the current exceeding the maximum metering range, Breakdown will be indicated
as follows. The red indicator light will illuminate on the RESET switch and an alarm will
sound. The display will show:
models 2925 and 2935
HI- F A I L MX XXX . X s
X . XX K V A C O F L mA
model 2945
W - H i g h MX XXX . X s
X . XX K V A C O F L mA
7. If the leakage current exceeds the metering range and neither a short circuit nor flash
over occurred, the red indicator light will illuminate on the RESET switch and an alarm
will sound. The display will show:
models 2925 and 2935
HI- F A I L MX XXX . X s
X . XX K V A C O F L mA
model 2945
W - H i g h MX XXX . X s
X . XX K V A C O F L mA