OPERATION
29
2. To set the test type and sequence for model 2945
On model 2945 only the Insulation Resistance (I) and Dielectric Withstand (W) test can
be run independently or as a sequence of tests. When “W” or “I” is selected, the
instrument executes only the test selected. When “W-I” is selected, first the Dielectric
Withstand test will execute and then the Insulation Resistance test will execute, selecting
“I-W” executes the two tests in the reverse order. To change the Test type on model 2945
only, please press the SET key until the display shows:
T e s t = W
or
T e s t = I
or
T e s t = W - I
or
T e s t = I - W
Press the Up or Down Arrow keys to toggle the test selection; then press the EXIT key to
exit to the test mode or toggle to another setting using the SET key.
Note: The Test mode type will affect the number of items displayed in the menu. The
menu will display only the items that are relative to the Test mode that has been selected.
3. To set the AC / DC mode for models 2935 and 2945
Press the SET key until the display shows:
Models 2935
M o d e = A C
or
M o d e = D C
Model 2945
W - M o d e = A C
or
W - M o d e = D C
Press the Up or Down Arrow keys to toggle the output function; AC or DC, then press the
EXIT key to exit to the test mode or toggle to another setting using the SET key.
4. To set the Output Test Voltage
Please press the SET key until the display shows:
Models 2925, 2935 Model 2945
V o l t a g e = X.XX KV
or
W - V o l t = X.XX KV
Please use the Up/Down Arrow keys to enter the desired test voltage, then press the EXIT
key to exit to the test mode or toggle to another setting using the SET key. The maximum
voltage which may be entered is 5000 volts AC or 6000 volts DC.
5. To set the High Leakage Current Limit