QUICK START
25
QUICK START
This quick start guide assumes the operator has some familiarity with automated hipot
testing and desires to use the "default" settings on the instrument. The default settings
shown will remain in memory unless you choose to override them with your own test
program. The default settings will appear in memories 1 through 5 for models 2925, 2935,
and 2945. The instrument default settings are as follows:
DEFAULTS
Input Voltage: 115 or 230 volts AC country specific
(rear panel switch selectable)
Lock: OFF
Test: W - I (2945)
Dielectric Withstand Test Mode
Voltage Output: 1,240 volts AC
High Current Trip: HI-LMT: 10 mA
Low Current Trip: LO-LMT: 0 mA (2925 and 2935)
Ramp Timer: 0 second (2925, 2935, 2945)
Dwell Timer: 1 second
Voltage Type: AC
AC Output Frequency: 60Hz
Continuity: OFF
Insulation Resistance Mode (model 2945)
Voltage Output: 1,000 volts DC
High Resistance Trip: I - High: 0 M
Low Resistance Trip: I - Low: 1 M
Delay Timer: 2 seconds