6. Close the drain valve (seen in Figure 7) on the bottom of the unit. Remove the fill port
plug on the front of the unit and fill the tank with 1.4 liters of distilled water. Replace the fill
port plug.
CAUTION: Tap water will increase earth leakage current. For proper system operation and
component life, use only DISTILLED WATER in the water reservoir.
7. Plug the unit into the appropriate electrical outlet. (See specifications for details. )