3. Specifications
IEC 61010-1 Over Voltage:
CAT II - 750Vac/1000Vdc
CAT III - 600Vac/dc
Pollution Degree 2
a. DCV
Range Resolution Accuracy Impedance
40mV 0.01mV
400mV 0.1mV
4V 0.001V ±(0.5% + 2 digits) 10MW
40V 0.01V
400V 0.1V
1000V 1V
b. ACV
Range Resolution Accuracy Impedance
40mV 0.01mV ±(1.5% + 5 digits)
400mV 0.1mV
4V 0.001V
40V 0.01V ±(0.75% +5 digits) 10MW
400V 0.1V
750V 1V ±(1.0% + 5 digits)
c. DCA
Range Resolution Accuracy Overload Protection
40µA 0.01µA ±(0.8% +2 digits)
400µA 0.1µA Fuse
4mA 0.001mA ±(0.8% +5 digits) 0.5Amp/600V
40mA 0.01mA
400mA 0.1mA
4A 0.001A ±(1.2% +5 digits) Fuse
10A 0.01A 10Amp/600V
9
UL61010-1
*Warning: Use only correct size, voltage and current rated fuses.
Test Leads: Use only correct type and overvoltage category rating.
d
d. ACA
Range Resolution Accuracy Overload Protection
400µA 0.1µA Fuse
4000µA 1µA ±(1.2% +5 digits) 0.5Amp/600V
40mA 0.01mA
400mA 0.1mA
4A 0.001A ±(1.5% +10 digits) Fuse
10A 0.01A 10Amp/600V
e. OHM (Resistance,
W
)
Range Resolution Accuracy Overload Protection
400W 0.1W ±(1.0% +5 digits) 600V DC or
4kW 0.001kW AC Peak
40kW 0.01kW ±(0.5% +3 digits)
400kW 0.1kW
4MW 0.001MW ±(1.0% +3 digits)
40MW 0.01MW ±(1.5% +10 digits)
g. Duty Cycle / Hz
Range
0.1 ~99.9% (0.5Hz to 500kHz, Width > 2uS)
Accuracy
((0.1% + 0.05% / kHz) +1 Count
h. Diode Test
Test Voltage Max Test Current Over Load Protection
2.7V Approx. 1mA 600 V DC or Peak AC
i. Continuity Buzzer
Test Voltage Threshold Over Load Protection
0.4 ~ 0.6V < 30W 600 V DC or Peak AC
UL61010-1
CAT II 750Vac/1000Vdc
CAT III 600Vac/dc