7. Calibration and adjustment
Calibration of gradient module
Calibration stage
based on position 0°
Place the GO! smart with the measurement surface horizontally
Briefly press buttons
and
Display -1- appears on the screen
Press the button briefly, display -1- flashes for approx. 5 seconds
Display -2- appears on the screen
Turn GO! smart again by 180°
Press the button briefly, display -2- flashes for approx. 5 seconds
Measurement display appears on the screen, calibration stage
1 is complete.
180°