A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-66
3.6.2.3.6 R. Follicle (Right Follicle) measurement
Function: To measure length and width of right follicle: R.FOL, R.FOW.
Measurement method:
1. Press the Calckey to pop up the Meas &
Calc menu, move the trackball, use the cursor
to select “GYN”, press theSETkey to go to
the submenu, move the trackball to select R.
Follicle, press theSETkey to go to R.
Follicle measurement mode.
2. Measure R.FO L of right follicle; see “Distance
Measurement” for the specific method.
3. Measure R.FO L of right follicle; see “Distance
Measurement” for the specific method.
4. The measurement result is displayed in the
result area.
5. Repeat steps (1) to (4) to go to the next R.
Follicle measurement.