EasyManua.ls Logo

Sony D-5 - Page 28

Sony D-5
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ye
CHI
djustment
-e
oscilloscope
IC
range)
+
(TP59)
»)
main
)
pages
‘emem-
side
e
refer-
ect
the
t
point
adjust
200nS.
n
con-
for
the
e
pulse
rection
@
Waveform
for
reference
VOLT/DIV:
1V
TIME/DIV:
5
uS
ov
Adjust
this
width
to
less
than
200nS.
6.
Connect
the
dual-trace
oscilloscope
as
in
step
1
and
confirm
that
the
triangular
wave
separation
or
overlap
amount
is
within
0.2
V.
@
Triangular
wave
reference
waveform
(shows
offset
voltage
off
in
one
direction)
VOLT/DIV:
50
mV
TIME/DIV:
50
uS
upper:
TP14
(1IC504
6),
GD)
lower:
TP15
(1C504
6),
@))
offset
voltage
Waveform
separation
or
overlap
amount
should
be
within
0.2
V.
B)
Simple
Method
D-5/D-50
D-5/D-50
|
7.
After
adjustment,
release service
mode.
(Refer
to
pages
9-11).
dua/-trace
oscilloscope
(DC
range)
main
board
en,
TP14
O———_©
TP15
O—S
era
=
1.
Connect
the
dual-trace
oscilloscope
to
main
board
test
points
TP14
and
TP1S.
2.
Put
the
set
into
service
mode
(refer
to
pages
9-11).
3.
Observe
the
oscilloscope
waveform
and
confirm
if
offset
voltage
is
off
on
the
+
or
side
relative
to
OV.
(Refer
to
triangular
wave
reference
waveform).
4.
Depending
on
the
offset
voltage,
adjust
RV503
so
that
the
tips
of
triangular
wave
are
0
V
as
fig.
@
and
@
below.
When
R548
is
mounted,
adjust
pattern
connec-
tion
of
R548
and
RV503
together.
@
Waveform
for
adjustment
®
When
offset
voltage
is
off
in
direction.
(
Adjust
that
the
upper
tips
of
waveform
are
0
V.)
|
B)
When
offset
voltage
is
off
in
+
direction.
(Adjust
that
the
lower
tips
of
waveform
are
OV.)
“VV
5.
After
adjustment,
release
service
mode
(See
page
9-11.)
Ov
@
Waveform
for
reference
(Offset voltage
is
off
in
direction)
upper:
TP14
(1C504
(6),
@))
VOLT/DIV:
100mV
lower:
TP15
(1C504
(6),
(8))
TIME/DIV:
10
uS
Check/
Adjustment
Location:
Main
board
3
of
R548
for
RV503
TP14
TPIS
R59
(GND)
TP57
TP58
when
offset
voltage
when
offset
voltage
is
off
in
+
direction
is
off
in
direction
4,
pattern
connection
PLL
FREE
RUN
FREQUENCY
CHECK
AND
ADJUSTMENT
Check/
Adjustment
Procedure
frequency
counter
=z
main
board
ooo
TP24
adjustment
o—-——_———-O
RV503
{PLCK)
Ly
TP14
.
Short
the
jumper
point
(A)
(PLL)
in
the
dia-
gram
below.
TP15
:
Connect
a
frequency
counter
to
main
board
test
point
TP24
(PLCK).
Put
the
set
into
service
mode
(see
page
9-11).
Check
that
the
frequency
counter
reading
is
4.3218
+0.04
MHz.
If
not,
adjust
L603
so
that
it
is
4.3218
+0.01
MHz.
After
adjustment,
release
service
mode
(see
page
9-11).
6.
Disconnect
the
jumper
point
shorted
in
step
1.
Check/
Adjustment
Location:
Main
board
L603
PLL
solder
jumper
point
an
DC:DC
converter
Short
for
checking
and
adijust-
ee
t
side
)
ment.
Disconnect
after
checking
and
adjustment.
PLL
solder
jumper
point
(PLCK)

Related product manuals