2. Furthermore all configurations shall comply with the
system standard IEC60601-1-1. Everybody who
connects additional equipment to the signal input
part or signal output part configures a medical
system, and is therefore, responsible that the system
complies with the requirements of the system
standard IEC60601-1-1.
If in doubt, consult the qualified service personnel.
3. The leakage current could increase when connected
to other equipment.
4. For this particular equipment, all accessory
equipment connected as noted above, must be
connected to mains via an additional isolation
transformer conforming with the construction
requirements of IEC60601-1 and providing at least
Basic Insulation.
5. This equipment generates, uses, and can radiate
radio frequency energy. If it is not installed and used
in accordance with the instruction manual, it may
cause interference to other equipment. If this unit
causes interference (which can be determined by
unplugging the power cord from the unit), try these
measures: Relocate the unit with respect to the
susceptible equipment. Plug this unit and the
susceptible equipment into different branch circuit.
Consult your dealer. (According to standard IEC60601-
1-2 and CISPR11, Class B, Group 1)
Important EMC notices for use in the medical
environments
The LMD-2765MD/LMD-2760MD needs special
precautions regarding EMC and needs to be installed
and put into service according to the EMC
information provided in this Instructions for Use.
The portable and mobile RF communications
equipment such as cellular phones can affect the
LMD-2765MD/LMD-2760MD.
Warning
The use of accessories and cables other than those
specified, with the exception of replacement parts sold
by Sony Corporation, may result in increased emissions
or decreased immunity of the LMD-2765MD/LMD-
2760MD.
Warning
If the LMD-2765MD/LMD-2760MD should be used
adjacent to or stacked with other equipment, it should be
observed to verify normal operation in the configuration
in which it will be used.
Guidance and manufacturer’s declaration-electromagnetic emissions
The LMD-2765MD/LMD-2760MD is intended for use in the electromagnetic environment specified below.
The customer or the user of the LMD-2765MD/LMD-2760MD should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The LMD-2765MD/LMD-2760MD uses RF energy only for
its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The LMD-2765MD/LMD-2760MD is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class D
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies