What to do if Sorvall Laboratory Equipment displays error E-39?
- LLindsey JohnsonSep 23, 2025
If your Sorvall Laboratory Equipment displays error code E-39, switch the instrument off and on again. If the error persists, contact KENDRO service.
What to do if Sorvall Laboratory Equipment displays error E-39?
If your Sorvall Laboratory Equipment displays error code E-39, switch the instrument off and on again. If the error persists, contact KENDRO service.
What to do if Sorvall Legend T is too noisy?
If your Sorvall Laboratory Equipment is making excessive noise, immediately press the "stop" key, or unplug the mains power cord in an emergency. Wait for the centrifuge to stop completely. Then, check if the rotor is loaded correctly and look for any broken tubes or damage to the rotor or motor that might be causing the noise.
What to do if Sorvall Legend T Laboratory Equipment displays message "bAl"?
If the message "bAl" appears on the display of your Sorvall Laboratory Equipment, open the instrument by pressing the "open lid" key. Then, check to ensure the rotor is properly loaded, and inspect for any broken tubes or rotor damage that may have triggered the imbalance switch. Also, verify that the trunnions of the swinging bucket rotor are properly lubricated.
What to do if Sorvall Legend T Laboratory Equipment displays message “Lift Lid“?
If the message "Lift Lid“ appears in the display of your Sorvall Laboratory Equipment, avoid placing any objects on the centrifuge lid and try lifting the lid slightly.
What to do if Sorvall Laboratory Equipment displays message "Lid"?
If the message "Lid" is flashing in the display of your Sorvall Laboratory Equipment, and the lid was opened manually during operation, switch the instrument off and then on again to continue the run.
What to do if Sorvall Laboratory Equipment displays error E-22 or E-23?
If your Sorvall Laboratory Equipment displays error code E-22 or E-23, switch the instrument off and on again. If the error persists, contact KENDRO service.
What to do if Sorvall Legend T displays error E-15 or E-16?
If your Sorvall Laboratory Equipment displays error code E-15 or E-16, switch the instrument off and on again. If the problem persists, contact KENDRO service.
What to do if Sorvall Laboratory Equipment displays error E-17?
If your Sorvall Laboratory Equipment displays error code E-17, switch the instrument off and on again. If the problem persists, contact KENDRO service.
What causes error E-21 on Sorvall Legend T Laboratory Equipment?
To address error E-21 on your Sorvall Laboratory Equipment, first check if a certified rotor is inserted. If the error occurred after a brief power failure, switch the instrument off and on again using the main switch.
What to do if Sorvall Legend T displays error E-33?
If your Sorvall Laboratory Equipment displays error code E-33, turn the instrument off and pull the main plug. Inspect and clean the ventilation slots if necessary. After about 60 minutes, you can restart the instrument, ensuring you observe the maximum permissible environmental temperature.
Guidelines for safe operation and warnings against misuse of the centrifuge.
Precautions for centrifuging toxic, radioactive, explosive, or flammable substances.
Rules for handling the centrifuge, including accessory use and lid operation.
Procedures for installing, handling, and maintaining various types of rotors.
Guidelines for maximum loading, filling tubes, and ensuring proper rotor balance.
Instructions for achieving aerosol-tight bio-containment during operation.
How to set speed, RCF, run time, temperature, and start the centrifugation process.
Guidelines for disinfecting the centrifuge, rotors, and accessories, especially after spills.
Procedure for manually opening the lid in case of power failure or malfunction.
A guide to identifying and resolving common operational errors and display messages.
| Maximum Speed | 15, 000 rpm |
|---|---|
| Max Speed | 15, 000 rpm |
| Maximum RCF | 21, 885 x g |
| Maximum Capacity | 4 x 100 mL |
| Temperature Range | -9°C to +40°C |
| Dimensions | 38.5 cm x 60.8 cm x 70 cm (H x W x D) |
| Weight | 82 kg |
| Type | Centrifuge |
| Capacity | 4 x 100 mL |
| Acceleration/Deceleration Profiles | 9 acceleration, 10 deceleration |
| Power Requirements | 50/60 Hz |