Table of Contents

Questions and Answers

  • L
    leviwaltonSep 23, 2025
    What to do if Sorvall Laboratory Equipment displays error E-39?
    • L
      Lindsey JohnsonSep 23, 2025
      If your Sorvall Laboratory Equipment displays error code E-39, switch the instrument off and on again. If the error persists, contact KENDRO service.
  • S
    Susan TranAug 25, 2025
    What to do if Sorvall Legend T is too noisy?
    • H
      Holly BrownAug 25, 2025
      If your Sorvall Laboratory Equipment is making excessive noise, immediately press the "stop" key, or unplug the mains power cord in an emergency. Wait for the centrifuge to stop completely. Then, check if the rotor is loaded correctly and look for any broken tubes or damage to the rotor or motor that might be causing the noise.
  • V
    Vanessa TravisAug 25, 2025
    What to do if Sorvall Legend T Laboratory Equipment displays message "bAl"?
    • L
      Lisa HughesAug 26, 2025
      If the message "bAl" appears on the display of your Sorvall Laboratory Equipment, open the instrument by pressing the "open lid" key. Then, check to ensure the rotor is properly loaded, and inspect for any broken tubes or rotor damage that may have triggered the imbalance switch. Also, verify that the trunnions of the swinging bucket rotor are properly lubricated.
  • D
    Dylan MosesAug 27, 2025
    What to do if Sorvall Legend T Laboratory Equipment displays message “Lift Lid“?
    • S
      sylviajonesAug 27, 2025
      If the message "Lift Lid“ appears in the display of your Sorvall Laboratory Equipment, avoid placing any objects on the centrifuge lid and try lifting the lid slightly.
  • S
    Shane MartinAug 27, 2025
    What to do if Sorvall Laboratory Equipment displays message "Lid"?
    • L
      Leah CarterAug 27, 2025
      If the message "Lid" is flashing in the display of your Sorvall Laboratory Equipment, and the lid was opened manually during operation, switch the instrument off and then on again to continue the run.
  • C
    Chad LawrenceSep 13, 2025
    What to do if Sorvall Laboratory Equipment displays error E-22 or E-23?
    • A
      annettefullerSep 14, 2025
      If your Sorvall Laboratory Equipment displays error code E-22 or E-23, switch the instrument off and on again. If the error persists, contact KENDRO service.
  • A
    Aaron PachecoSep 12, 2025
    What to do if Sorvall Legend T displays error E-15 or E-16?
    • D
      Douglas BrownSep 12, 2025
      If your Sorvall Laboratory Equipment displays error code E-15 or E-16, switch the instrument off and on again. If the problem persists, contact KENDRO service.
  • M
    Marcus MendozaSep 12, 2025
    What to do if Sorvall Laboratory Equipment displays error E-17?
    • H
      Heather RodgersSep 12, 2025
      If your Sorvall Laboratory Equipment displays error code E-17, switch the instrument off and on again. If the problem persists, contact KENDRO service.
  • C
    crystal10Sep 4, 2025
    What causes error E-21 on Sorvall Legend T Laboratory Equipment?
    • O
      omaddoxSep 4, 2025
      To address error E-21 on your Sorvall Laboratory Equipment, first check if a certified rotor is inserted. If the error occurred after a brief power failure, switch the instrument off and on again using the main switch.
  • A
    Amanda StephensonSep 2, 2025
    What to do if Sorvall Legend T displays error E-33?
    • H
      hayskristinSep 2, 2025
      If your Sorvall Laboratory Equipment displays error code E-33, turn the instrument off and pull the main plug. Inspect and clean the ventilation slots if necessary. After about 60 minutes, you can restart the instrument, ensuring you observe the maximum permissible environmental temperature.

Summary

Control Panel Overview

General Safety Instructions

Proper and Improper Use

Guidelines for safe operation and warnings against misuse of the centrifuge.

Handling Hazardous Materials

Precautions for centrifuging toxic, radioactive, explosive, or flammable substances.

Safe Centrifuge Handling

Rules for handling the centrifuge, including accessory use and lid operation.

Accessories and Rotors

Before Use: Installation and Connection

Operation Procedures

Installing and Handling Rotors

Procedures for installing, handling, and maintaining various types of rotors.

Loading the Centrifuge Rotor

Guidelines for maximum loading, filling tubes, and ensuring proper rotor balance.

Aerosol-Tight Operation

Instructions for achieving aerosol-tight bio-containment during operation.

Setting Parameters and Starting Run

How to set speed, RCF, run time, temperature, and start the centrifugation process.

Maintenance and Care

Disinfection Procedures

Guidelines for disinfecting the centrifuge, rotors, and accessories, especially after spills.

Troubleshooting Guide

Emergency Lid Release

Procedure for manually opening the lid in case of power failure or malfunction.

Common Error Symptoms and Solutions

A guide to identifying and resolving common operational errors and display messages.

Technical Data and Specifications

Sorvall Legend T Specifications

General IconGeneral
Maximum Speed15, 000 rpm
Max Speed15, 000 rpm
Maximum RCF21, 885 x g
Maximum Capacity4 x 100 mL
Temperature Range-9°C to +40°C
Dimensions38.5 cm x 60.8 cm x 70 cm (H x W x D)
Weight82 kg
TypeCentrifuge
Capacity4 x 100 mL
Acceleration/Deceleration Profiles9 acceleration, 10 deceleration
Power Requirements50/60 Hz