RT6000
and
T6000
Centrifuges
SORVALL®
Centrifuges
7-2. Calibration
This
section
contains
check and adjustment
procedures
for
calibrating specific
circuit
functions in
order
to maintain
the
best performance
of the
centrifuge.
These
calibrations
should
only
be
performed
when
the
replacement
of
centrifuge
components
or
system
aging
makes adjustments
necessary.
theproHcedures.
a.
Temperature
Meter Adjustment
(RT6000
ONLY)
With
the
POWER
switch
set to OFF, turn
the
adjustment screw
(using a
screwdriver)
until
the
pointer reads
-20°C
(see
figure 7-5).
ADJUSTMENT
SCREW
2/81
RT-6000'6
Figure
7-5.
Temperature
Meter
Adjustment
b.
Tachometer Zero Adjustment
With the
POWER
switch
set
to
OFF, turn
the
adjustment screw
(using a
screwdriver)
until the
pointer reads
0
(see
figure
7-6).
7-22