SORVALL®
Centrifuges
Introduction
&
Description
1-3.
Centrifuge Specifications
Maximum
Speed*
..........................6000
rpm**
Operating
Temperature
Range
(Refrigerated
centrifuges
only)
..........................
.-5°C
to
+25°C
Maximum
heat
output
during
operation
............................
1.76
kW
(6000 Btu/h)—Refrigerated
centrifuges
0.88
kW
(3000 Btu/h)— Non-refriger-
ated
centrifuges
Electrical Requirements:
Refrigerated centrifuges
........................115
Vac,
60
Hz,
15
A,
single
phase
220
Vac,
50
Hz,
8
A,
single
phase
Non-refrigerated centrifuges
....................115
Vac,
60
Hz,
8
A,
single
phase
220
Vac,
50
Hz,
4
A,
single phase
Power Receptacles:
115
Vac Receptacle
...........................
NEMA 5-15R
or equivalent
220
Vac
Receptacle
...........................NEMA
6-15R
or
equivalent
Dimensions:
Width
......................................81
cm
(32
inches)
(T6000D and T6000B,
SN
9101800
and above
only)
.................66
cm
(26
inches)
Depth
......................................64
cm
(25
inches)
Height
......................................32
cm (12.5 inches)
Mass (Weight):
...............................85
kg
(187
Ib)
(T6000D and
T6000B,
SN 9101800
and above
only)
.................78
kg
(160
Ib)
"
Maximum
speed
is dependent on rotor
and
line
voltage.
"Speed
in
revolutions
per
minute
(rpm)
is related
to
angular velocity,
<u,
according to
the following:
<a
=
(^(il)"
(rpm)
(0.10472)
Where
co = rad/s.
All further
references
in this
manual
to
speed
will
be designated
as
rpm.