10
7030-0068-26
28Jul16
Customer Service, Canada/US (719) 633-8333 / (800) 231-0978
Customer Service, Europe +31 33 434 7050
CVX-300
®
Excimer Laser System Version 26 Operator’s Manual
English / English
Specications
The excimer laser is a pulsed laser with the following nominal specications.
Active Medium XeCl
Wavelength 308 nm
Catheter Output Fluence* 30 - 80 mJ/mm
2
Repetition Rate Range* 25 - 80 Hz
Pulse Width 125-200 ns, FWHM
Weight 650 lbs / 295 kg
Length 49 in / 125 cm
Height 35 in / 89 cm - unit
7-9 in / 18-23 cm - control panel
Width 25 in / 62 cm
(All Dimensions Approximate)
Power Requirements 200 – 230 V ~ - Single Phase
50/60Hz
16 Amp
Environmental Specications
- Operating temperature: 12°C to 30°C (54°F to 86°F)
- Storage temperature: 0°C to 60°C (32°F to 140°F)
- Operating humidity: 5 to 95% relative humidity, non-condensing
- Storage humidity: 5 to 95% relative humidity, non-condensing
System footswitch is IPX8 rated.
Accessories: power cord, footswitch, interlock plug, safety glasses, and Spectranetics
approved ber optic catheters.
Environment: Avoid exposing the laser system to extreme temperatures (below 32°F or 0°C,
above 140°F or 60°C).
* Dependent on ber optic catheter in use and the CVX‑300 software installed; see
the Instructions for Use documentation supplied with each ber optic catheter for
specic information.