28
7030-0068-26
28Jul16
Customer Service, Canada/US (719) 633-8333 / (800) 231-0978
Customer Service, Europe +31 33 434 7050
CVX-300
®
Excimer Laser System Version 26 Operator’s Manual
English / English
Reset: The Reset button is used in combination with the Standby button, Pulses Delivered
and Treatment Time buttons.
Standby: If the system is in either Ready or Calibrate mode, depressing the Standby
buttons places the Excimer Laser System in Standby mode. The Standby button and the
green status light are illuminated.
Keyswitch: The keyswitch is the power control for the Excimer Laser System.
In case of emergency, the Excimer Laser System can be powered down by depressing
the Emergency Shuto located on the back panel. Caution should be taken not
to activate the Emergency Button accidentally. To reactivate the system, rotate
the Emergency Shuto clockwise until it ascends and turn the keyswitch to the
OFF( ) position and then to the ON ( | ) position. Refer to the Reset and Standby and
Warning below to bypass the CVX-300 Excimer Laser warm-up.
Treatment Time: Depressing the Treatment Time button enables the total lasing time of
the procedure to be visible in the display window.
Reset and Pulses Delivered: The Reset and Pulses Delivered buttons are depressed
simultaneously to reset the total pulses delivered during the procedure.
Reset and Standby: The Standby and Reset buttons are depressed simultaneously to
bypass the ve minute warm-up phase if the Excimer Laser System parameters, uence and
rate, are set. The CVX-300 must not be o for longer than 30 seconds in order to use this
feature. See Warning below.
Reset and Treatment Time: The Reset and Treatment Time buttons are depressed
simultaneously to reset the total lasing time of the procedure.
Warning
Bypassing the warm‑up period may damage components in
the Excimer Laser System if the system has been o for longer
than 30 seconds.

Questions and Answers: