Section 2 -Installation & Operating Procedures
Setup/Installation
Environment
Make sure that your spectrophotometer is placed as far
as possible from any strong magnetic or electrical
fields, or any electrical apparatus that may generate
high-frequency fields.
1. Select the appropriate glassware and adapters. For
more information about glassware, adapters and cell
holders, refer to Table 4-1 on page 4-1 and Table
4-2 on page 4-2.
2. If you are connecting the instrument to an Accessory
Printer, computer or chart recorder, refer to Table
2-1 for information about cabling requirements.
The instrument should be installed in an area that is
free of dust, corrosive gases and severe vibrations.
In addition, there should be no obstructions that could
hinder the flow of air under and around the instrument.
Table 2-1 Cabling requirements for SPECTRON1C1> 20 series of spectrophotometers
200+
20D
DEVICE
~
20/20+
.333174 -Patch Cord
.333174 -Patch Cord
Analog
Recorder
.333174 -Patch Cord
.333132 -Serial Interface
Cable Kit
Not available
Accessory
Printer
.333177 -Serial Interface
Cable
.335488-1601 (included
with Serial Printer 3~)
.333132 -Serial Interface
Cable Kit
Not available
ffiMPC
XT
.333177 -Serial Interface
Cable
.335243 -Null Modem
Cable
OR
.333132 -Serial Interface
Cable Kit
.333132 -Serial Interface
Cable Kit
ffiMPC
AT
Not available
.333177 -Serial Interface
Cable
.335243 -Null Modem
Cable
.345002-111 -Adapter
Cable
OR
.333132 -Serial Interface
Cable Kit