EasyManuals Logo

Sperry instrument DM-4100A User Manual

Sperry instrument DM-4100A
23 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #19 background imageLoading...
Page #19 background image
5. Carefully inspect the other DCV ranges. Your readings should
be within specification ±0.5% + 1 digit.
6. There is no adjustment for ACV. Calibrate DCV first.
7. Carefully inspect the ACV ranges. Your readings should be
within ± 1.2% + 4 digits of the ACV calibration source.
8. Set the output of the DC calibrator for 1.9 ±0.02% and connect
it to the "10A" and "COM" input terminals.
9. Adjust R001 (shunt resistor) until the display reads 1.9A.
10. If the reading is over 1.9A, add solder to R001. If the reading
is under 1.9A, shave away lightly some of the solder and metal
from R001.
11. Carefully inspect the other DCA ranges. Your readings should
be within specification ±1.0% + 1 digit.
12. Turn off calibrator and disconnect from the DMM.
13. Install the back case and insert the back case screw.
17
DM4100 04_06 5/3/06 1:09 PM Page 20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sperry instrument DM-4100A and is the answer not in the manual?

Sperry instrument DM-4100A Specifications

General IconGeneral
BrandSperry instrument
ModelDM-4100A
CategoryMultimeter
LanguageEnglish

Related product manuals