T14 Patient and Staff Badges
User Guide 17
Battery Levels and Charging
The badges have the following 3 (three) battery levels that are shown in
MobileView:
• High – Indicates the badge has more than 30% battery capacity
• Medium – Indicates the badge has between 10-30% battery capacity
• Low – Indicates the badge has less than 10% battery capacity
Note
It is recommended to charge the badges during the Medium level or
after a week of use, whichever comes first.
Badges are recharged using the T14 Badge Charging Station. Please refer to the
T14 Badge Charging Station User Guide.
Battery Capacity
The Badge’s battery capacity degrades after a period of time, depending on the
number of charge cycles (see table below). A charge cycle means using all the
battery's capacity.
Number of Charges Badge Battery Capacity
300 80%
400 75%
500 70%
600 65%
The reduction in capacity is much slower when the battery is partly discharged.
For T14 Badges, the battery is partly discharged if the badge is used for less than
a week and is recharged again.
When the operating time of a charged battery is significantly reduced, the
badge should be disposed of according to facility procedures in your
jurisdiction.