Table 3
The TAG1410 is intended for use in the electromagnetic environment specified below. The customer or the
user of the TAG1410 should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance level
Electromagnetic environment
guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC 61000-4-3
Not Applicable
10 V/m
80 MHz to 2,7 GHz
Not Applicable
10 V/m
80 MHz to 2,7
GHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the TAG1410, including cables, than
the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
𝑑 =
12
10
𝑃
= 1.2
𝑃 80 MHz to 800 MHz
𝑑 =
23
10
𝑃
= 2.3
𝑃 800 MHz to 2.7 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m)
a
.
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey,
b
should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.