T14 Patient and Staff Badges
User Guide 26
communications equipment could cause interference if it is inadvertently brought into
patient areas.
NOTE 3 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Table Immunity to RF Wireless Communication Equipment
Test
Frequency
Band Service Modulation
Maximum
Power
Distance
Immunity
Test Level
Compliance
Level
(MHz)
(MHz)
(W)
(m)
(V/m)
(V/m)
385
380-
390
TETRA 400
Pulse
Modulation
18 Hz
1.8 0.3 27 27
450
430-
470
GMRS 460,
FRS 460
FM
± 5 kHz
Deviation
1 kHz sine
2 0.3 28 28
710
704-
787
LTE Band
13, 17
Pulse
Modulation
217 Hz
0.2 0.3 9 9
745
780
810
800-
960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
Modulation
18 Hz
2 0.3 28 28
870
930
1720
1700-
1990
GSM 1800;
CDMA
1900;
GSM 1900;
DECT;
LTE Band 1,
3,
4, 25; UMTS
Pulse
Modulation
217 Hz
2 0.3 28 28
1845
1970
2450
2400-
2570
Bluetooth,
WLAN,
802.11
b/g/n,
RFID 2450,
LTE Band 7
Pulse
Modulation
217 Hz
2 0.3 28 28