Page 12
Fill Timer
The LN2 SUPPLY LED will light and a number will be displayed. This is the Fill Timer setting in minutes.
Use the TEST key to increase the Fill Timer setting.
Use the MUTE key to decrease the Fill Timer setting.
The Fill Time is adjustable in 5 minute steps between 0 and 125 minutes.
If the solenoid valve is operated for longer than this pre-set time an alarm condition will occur.
Press the FILL key when finished.
Remote Alarm Timer
The REMOTE LED will light and a number will be displayed.
This is the Remote Alarm Timer setting in minutes.
Use the TEST key to increase the Remote Alarm Timer setting.
Use the MUTE key to decrease the Remote Alarm Timer setting.
The Remote Alarm Timer is adjustable in 5 minute steps between 0 and 60 minutes.
The Remote Alarm Timer sets the time which must elapse between the first alarm condition and activation of
the remote alarm relay. The exception to this is the Overfill alarm which defaults to 5 minutes irrespective of the
Remote Timer setting.
Press the FILL key when finished.
Lid Timer
The LID LED will light and a number will be displayed. This is the Lid Timer setting in minutes.
Use the TEST key to increase the Lid Timer setting.
Use the MUTE key to decrease the Lid Timer setting.
The Lid Timer is adjustable in 5 minute steps between 0 and 30 minutes.