Strasbaugh Software Calibration Procedures
Version 4.0 - February 1998 10 - 157
7. When the goal is met, record the value that met the goal.
Example: 1200
8. At the Calibration/Configuration Menu, enter the new calibration value
for the variable 38, Spindles down soft value.
Example: 1200
C. CHECK AND ADJUST VARIABLE 39, THE SPINDLES DOWN
HARD DOWN VALUE
Repeat as for variable 38, substituting 39 for 38 and making other substitutions
shown in the text of the procedure as is appropriate.
D. CHECK AND ADJUST VARIABLE 48, THE SPINDLES DOWN
EXTRA SOFT VALUE
Repeat as for variable 38, substituting 48 for 38 and making other substitutions
shown in the text of the procedure as is appropriate.