EN-30
Instructions
Point of Use
Wipe excess soil from the device using disposable paper towels.
If an automated reprocessing method will be used, rinse any
channels in the device with 50mL of sterile distilled water
immediately after use.
Containment and Transportation
Reprocess the device as soon as reasonably practical following use.
Transport the device in a tray to avoid damage.
Preparation for Cleaning
1. Disassemble the coupler from the scope and camera head.
2. Prepare an enzymatic detergent according to the manufacturers
recommendations (one ounce per gallon of tap water at 35 - 40°C).
3. Wipe the entire device with the detergent, using a clean cloth.
4. Immerse the device in the detergent. Inject any inside regions of the
device with 50mL of the detergent solution to remove loose debris.
5. Soak the device in the detergent for at least 15 minutes.