EN-34
Steam
• Steam sterilization is intended only for camera heads and couplers
marked
AUTOCLAVE.
• Rapid cooling, or “quenching,” the coupler after autoclaving will
result in product damage.
• The water used in the autoclave process must meet standards for
clean steam per AAMI ST 79 Appendix M – Steam Quality
requirements.
Ethylene Oxide (EO)
Note for United States users: For all autoclave-compatible devices, Stryker
recommends using steam sterilization instead of liquid chemical sterilization.
Sterilizer Type “Flash”
Gravity
“Flash”
Prevacuum
Gravity Prevacuum
Minimum
Temp erat ure
132 - 137°C
(270 - 279°F)
132 - 137°C
(270 - 279°F)
132 - 137°C
(270 - 279°F)
132 - 137°C
(270 - 279°F)
Minimum
Cycle Time
10 minutes 3 minutes 10 minutes 3 minutes
Product
Configuration
Unwrapped Unwrapped Double
wrapped
Double
wrapped
Drying Time — — 60 minutes 60 minutes
Preconditioning parameters
Temperature 55 ± 2°C (131 ± 4°F)
Exposure
Concentration (100% EO) 725 mg/L
Temperature 131 °F (55°C) ± 5 °F
Time 1 hour
Chamber Humidity 70% RH (50-80 %) ± 5%
Aeration parameters
Aeration Time 12 hours
Temperature 35 – 55°C (95° - 131°F)