18 www.stryker.com
EN 7205-001-700 Rev-AA
Specifications (continued)
Guidance and manufacturer’s declaration - electromagnetic immunity
The System 7 Dual Trigger Rotary Handpiece (REF 7205-000-000) is intended for use in the electromagnetic
environment specified below. The customer or the user of the System 7 Dual Trigger Rotary Handpiece (REF
7205-000-000) should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 V 0.15 MHz –
80 MHz
6 V in ISM bands
between 0.15 MHz
and 80 MHz
80% AM at 1 kHz
N/A IEC 60601-1-2 3rd Edition:
Portable and mobile RF equipment should be used
no closer to any part of the System 7 Dual Trigger
Rotary Handpiece (REF 7205-000-000), including
cables, than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
d=1.2√P 80 MHz to 800 MHz
d=2.3√P 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Radiated RF
IEC 61000-4-3
b
3 V/m
80 MHz to 2.7
GHz
80% AM at 1 kHz
27 V/m
385 MHz, pulse
modulation 18 Hz,
Maximum power
= 1.8 W
28 V/m
450 MHz, FM ±
5 kHz deviation,
1 kHz sine,
Maximum power
= 2 W
b
3 V/m
80 MHz to 2.7
GHz
80% AM at 1 kHz
27 V/m
385 MHz, pulse
modulation 18 Hz,
Maximum power
= 1.8 W
28 V/m
450 MHz, FM ±
5 kHz deviation,
1 kHz sine,
Maximum power
= 2 W