OPERATION MANUAL Sumitomo Heavy Industries, Ltd.
SRDK Series CRYOCOOLER 19
2-4 SITE REQUIREMENT
2-4-1 REQUIRED FOR AIR COOLED COMPRESSOR UNIT
・An almost level and even area in the field will be selected to install the Compressor Unit.
・An area to be influenced by splashing water and/or dusts will not be selected to install the Compressor
Unit.
・A clean environmental condition without dirt and/or free from an exhausted heat will be selected to install
the Compressor Unit in the field.
・An efficient ventilated area will be required to free from an exhausted heat of the Compressor Unit in the
field.
・A suitable air conditioning capacity will be secured for an installing area for the Compressor Unit in the
field.
・Any object and/or obstacle cannot be positioned on a ventilation fan outlet in a top area of the enclosure
and/or on surroundings of the Compressor Cooler.
・Any heat sensitive object cannot be positioned on surroundings of the Compressor Unit.
AMBIENT TEMPERATURE CONDITION
The ambient temperature must be between 5 deg.C (41 deg.F) and 28 deg.C (82.4 deg.F) to get the
specified cooling capacity. The system can operate up to 35 deg.C (95 deg.F) with less than 5%
cooling capacity down. The maximum relative air humidity is 85%RH.
HELIUM SUPPLY SYSTEM
A helium supply system is necessary if you need to decontaminate the helium gas, or or charge the
helium gas that has leaked out of the system. A helium supply system includes a Grade 5 (99.999%
up pure) helium gas bottle, a regulator, an outlet valve, and a charging hose or equivalent delivery line.
POWER SOURCE
Ensure the correct AC power source is available for the Compressor Unit. See “TECHNICAL
INSTRUCTION” of CSA-71A, for AC power source requirement.
SAFETY / SEISMIC REQUIREMENT
Secure to lock the locking device of compressor castor.
SERVICE AREA
Air-cooled Compressor Unit should have enough space for air-flow as shown in “TECHNICAL
INSTRUCTION” of CSA-71A.