OPERATION MANUAL Sumitomo Heavy Industries, Ltd.
SRDK Series CRYOCOOLER 27
2-7 CONNECTING FLEX LINES
CONNECTING PROCEDURE (in case of using the “Buffer Tank”)
In case of using the “Buffer Tank”, insert the “Buffer Tank” between the Compressor Unit and Flex Lines.
Refer to the Figure 2.1, outline view of the “Buffer Tank”, and the Figure 2.2, “Buffer Tank”
interconnecting diagram.
Connect the “Buffer Tank” with following procedure.
1) Connect the female coupling of “Tee-Adapter” to the male connector of the Compressor Unit for both
“Supply” and “Return”.
2) Place the “Buffer Tank” at the right side of the Compressor Unit.
3) Connect the “Tee-Adapter” and “Buffer Tank” by two “8A x 1m Flex Line”.
4) Connect the Flex Lines to the Tee-Adapters. Do not miss-match the “Supply” and “Return” marking
for both Flex Lines and Compressor Unit