OPERATION MANUAL Sumitomo Heavy Industries, Ltd.
SRDK Series CRYOCOOLER 54
6 TROUBLE SHOOTING FLOW CHART
FLOW C
Cold Head Cooling Power is low
Check the leakage of
the flex line connections
Abnormal
Correct the error(s)
then
Adjust the filling pressure
Normal
No intake and exhaust noise
from Cold Head
Check the ambient temperature
Control the circumstance to
the specified temperature
Abnormal
Normal
Check the Cold Head
Power Cable connection
Correct the error(s)
Abnormal
Normal
Cold Head need to be rapaired
(Contact to SHI)
Check the Cold Head
Power Cable snapping
Normal
Abnormal
Cold Head Power Cable
need to be replaced
Check the fuses of
the Compressor Unit
Abnormal
Replace the fuse(s)
Normal
Normal
Check the filling pressure Abnormal
Adjust the filling pressure
Normal
<For CSW-71C,D>
Check the cooling water
(temperature, flow rate, etc.)
Abnormal
Correct the error(s)
Normal
< For CSA-71A >
Check the air flow space of
the Compressor Unit
Abnormal
Ensure the enough space
Check the termal interface of
the Cold Head 1st and 2nd stage
Abnormal
Repair the interface
for good thermal conductivity
Normal
Check the vacuum
Correct the leak and fix it
Abnormal
Normal
Check the heat load to
the Cold Head
Correct the error(s)
Abnormal
Normal
Cold Head need to be rapaired
(Contact to SHI)
Following infomation will be useful
for failure mode analysis
Damage of Cold Head Cilynder:
Warm up and remove the Cold Head
and check the dent, bend or damage of
the Cold Head Cilynder.
Reference;
Compressor Unit
Technical Insuruction
Section 1-1
Reference;
Compressor Unit
Technical Insuruction
Section 1-1
Reference;
Operation Manual
Section 4-1-1
Reference;
Operation Manual
Section 4-1-1, 5-4
Reference;
Operation Manual
Section 4-2
Reference;
Operation Manual
Section 4-1-1, 5-4
Reference;
CSA-71A
Technical Insuruction
Section 2-1
Reference;
CSA-71A
Technical Insuruction
Section 2-1
< For CSA-71A >
Check the dust of
heat exchanger
Abnormal
Clean the heat exchanger
Reference;
CSA-71A
Technical Insuruction
Section 3-1-2
Reference;
CSA-71A
Technical Insuruction
Section 3-1-2
Normal
Reference;
Cold Head
Technical Insuruction
Section 1-1
Reference;
CSW-71C, CSW-71D
Technical Insuruction
Section 1-1
Reference;
CSW-71C, CSW-71D
Technical Insuruction
Section 1-1
Reference;
Operation Manual
Section 2-8
Reference;
Operation Manual
Section 2-8
Reference;
Compressor Unit
Technical Insuruction
Section 3-2
Reference;
Compressor Unit
Technical Insuruction
Section 3-2

Table of Contents

Related product manuals