18
Step 7: Insert the Tension Rod
The tension rod is inserted into the
hole on the table frame and then
screwed into the white drop block. Be
careful not to cross thread the hole
when screwing the rod into the drop
block (it should turn in smoothly).
Turn the rod just until you feel firm
resistance then stop.
Problem Solution
Bolt spins or wont tighten Thread has been stripped. Call support .
Drop wont cock after assem-
bly.
Some surface rust might have formed in shipment. Easy solution is to rap-
idly and gently move the cocking arm up and down until the drop does
cock. We then suggest spraying a very small amount of WD40 on the
plunger to lubricate it.
With time the drop seems stiff Spray the drop with a small amount of WD40.
Feel a sharp object through a
cushion
Every effort has been made to ensure that all of the cushion attachment
screws were not over driven into the cushion. Should you think that you
feel a sharp object through a cushion, look under the table at the point
where you feel the object. If you see an attachment screw, use a Phillips
head screw driver to back the screw out several turns. This should solve
the problem.
Drop wont cock after pro-
longed use.
Spray between the lifting arm and the white block some WD 40. Then us-
ing a quick tapping method, move the cocking rod up and down until the
drop seats itself and returns to normal function.
We are always available to help! Contact us via phone, email or chat on our website
(sunchirotable.com) at any time and we will be happy to assist you!
Troubleshooting