GE-Marquette MAC 5000/5500, Tango M2 Interface Notes Page 1 of 5
Marquette: MAC 5000, 5500
Tango M2 Interface Notes
To setup Tango M2 with the MAC 5000/5500 stress system, simply follow the directions below.
1. Verify Correct RS-232, ECG Trigger, and Splitter Cables
RS-232 Cable used to communicate with the MAC 5000/5500. This cable enables the MAC
5000/5500 to prompt Tango M2 when it needs a BP measurement, and allows the Tango M2 BP
reading to be transferred to the MAC 5000/5500 display and reports.
o SunTech Part Number = 91-0010-01
o Connection Ends:
ECG Trigger Cable provides the ECG signal from the MAC 5000/5500 to the Tango M2.
o SunTech Part Number = 91-0009-01
o Connection Ends:
MAC 5000/5500 Splitter Cable
Connection
8 pin mini DIN male
Tango M2 Connection
9 pin female
MAC 5000/5500 Connection
9 pin male
Tango M2 Connection
BNC male

Other manuals for SunTech TANGO M2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SunTech TANGO M2 and is the answer not in the manual?

SunTech TANGO M2 Specifications

General IconGeneral
Measurement MethodOscillometric
Power Source4 AA batteries or AC adapter
ConnectivityUSB
DisplayLCD
TypeUpper Arm
UseProfessional
Measurement SiteUpper Arm
Accuracy (Pressure)±3 mmHg
MemoryUp to 99 readings
Device TypeBlood Pressure Monitor
Cuff Size22-32 cm (Adult)
Measurement Range (Pressure)0 to 300 mmHg
Measurement Range (Pulse)40-180 beats per minute
Accuracy (Pulse)±5%

Related product manuals