Perform a One Plane registration before making an internal markers
registration.
( 7"=™7
This is the easiest method to automatically register the 2nd Modality, with
the use of an omniTRAX™ patient tracker, CT or MR, see omniTRAX™
and omniTRAX™ MR sections for details.
1. Click on the omniTRAX™ tab
2. Click on one of the four points of the omniTRAX™ markers seen
on the 3D view (make sure the auto detection of the omniTRAX™
markers in enabled)
3. The 3 other points will be automatically detected and selected.
If the auto detection fails, you can select the markers manually, either
on the 3D view or on one of the 3 MPR views, which can be scrolled
using the TouchRing or PW.
The Settings option Optimize marker position during click
allows you to ease the manual markers selection workflow.
4.
Click on Register Markers
The 2 modalities are now registered.