RMK-700 Operating Manual
10
properly, which indicates that the instrument can be measured normally.
2.3 Measurement
Measure the Model eye before measuring the patient
Measurement Procedure
(1) Turn ON the power
(2) Set the model eye on
Remove the model eye and align the holes of the model eye with the
the chin rest, and insert pins.
(3) Release stage lock
Loose the stage solid bolt downward.
(4) Adjust height of Test Model Eye
Adjust the height of the chin rest by rotating Height Adjustment Knob
so that The Test Mode Eye is aligned with the Height Adjustment Mark on
the headrest.
(5) Enter KR or REF Mode
Press MODE button until either one is displayed.
(6) Adjust the position and focus on the Model Eye.
Note: The Model eye is developed by the National Institute of Metrology, so the
customers configure it according to their own needs.
Height adjustment: rotate Joystick and Height Lifting Knob of chin rest
to adjust height;
Focus adjustment: Move the joystick all around, and focus when the
center becomes a cross.
(1) Measurement
Manual measurement: Press the measurement button. If measurement is