EasyManua.ls Logo

Sylvac S Dial WORK ANALOG NANO Smart - Dynamic Measuring Mode; Instrument Adjustment for Dynamic Mode; Performing Dynamic Measurements

Sylvac S Dial WORK ANALOG NANO Smart
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SET
18
E
11. Adjustment and use of IntE dynamic measuring mode
Certain applications need to adjust the instrument to the MIN (or MAX) measured value. In this case, proceed as follows :
11.1 Adjustment of the instrument
- Enter a Preset value corresponding to the actual size of the standard (see chap. 3)
- In the IntE menu, select the mode -AMIN (automatic selection of DIR- and mode MIN) or+AMIN (automatic
selection of DIR+ and mode MIN) or +AMAX (automatic selection of DIR+ and mode MAX) depending on application
- Make a standard measurement (going through the turnaround point)
- Adjust the instrument by selecting the SEt mode and pressing on the button (see chap. 3)
- The instrument is adjusted and ready to measure.
11.2 Measure
- Make the measurements going through the turnaround point. The digital display stores and displays the MIN (or
MAX) measured value.
- Before each new measurement, reset the measured value by a short press on