WARNINGS AND CAUTIONS FOR SAFE USE
Failure to follow all of the warnings and cautions listed below may result in
damage to the instrument and/or serious personal injury. Tanaka Scientific
Limited will not assume any responsibility for faults or accidents arising from
failure to adhere to these instructions.
Installation Environment
!
Warning: The instrument must be installed on a flat, sturdy surface
to prevent the instrument from falling over.
Coolant Circulation Piping
!
Caution: Ensure that each joint in the coolant circulation piping is
securely fastened to prevent coolant leakage, as this may
result in instrument failure.
!
Caution: Do not use methanol as cooling medium. O-ring of
solenoid valve may have damage and the unit will not work
correctly.
70% of ethylene glycol and 30% of distilled water mixture
will be the best for this unit.
Electrical Connection
!
Warning: Observe the following instructions when handling power
cables and electric connections:
・ Do not connect or disconnect any cables or electrical
connections when the mains power is ON.
・ Do not handle any cables or electrical connections with
wet hands.
・ Be sure to use the plug to remove cables. Never use
the cord to pull a plug from an outlet.
・ Do not place heavy items on cords or cable.