SPECIFICATION
TE-10A
Operating temperature range -20 to +95°C
Temperature range without cooling mechanism Ambient +5 to 95°C
Temperature selection Analogue
Temperature display Analogue scale
Temperature stability ±0.01°C
Set point accuracy ±2°C
Method of control Proportional
Temperature sensor Thermistor
Nominal heater power 230V 1000W
120V 1000W
100V 850W
Maximum watts density 6 W/cm
2
Pump capacity Maximum flow 10 l/min
Maximum pressure 145 mbar
The TE-10A has passed:
EN 50081-1:1992 Electromagnetic Compatibility; Generic emission standard.
EN 50082-1:1992 Electromagnetic Compatibility; Generic immunity standard (Performance criterion B).
The specification was achieved in an 8 litre bath with a ball blanket according to DIN 58966.
Protection against hazards IP30
Safety device classification 1W
24