SPECIFICATION
TU-20D
Operating temperature range -40 to +200°C
Temperature range without cooling mechanism Ambient +5 to 200°C
Temperature selection Digital
Temperature display Digital LED
Temperature stability ±0.005°C
Set point accuracy ±1°C
Method of control PID
Temperature sensor PRT
Nominal heater power 230V 1800W
120V 1500W
100V 1250W
Maximum watts density 6.2 W/cm
2
Pump capacity Maximum flow 10 l/min
Maximum pressure 145 mbar
Communications RS232
The TU-20D has passed
EN 50081-1:1992 Electromagnetic Compatibility; Generic emission standard.
EN 50082-1:1992 Electromagnetic Compatibility; Generic immunity standard (Performance criterion B).
The specification was achieved in an 8 litre bath with a ball blanket according to DIN 58966.
Protection against hazards IP30
Safety device cassification 2